TEMI880
No. | Button | Function | |
1 | Main | Change the page. | |
2 | General execute button | It is used general execution orselection. | |
3 | Page Up/Down button | It is used page conversion on thesame screen. | |
4 | Next page button | It is used to convert to next page. | |
5 | Setting value select button | It is used setting value selection. | |
6 | Setting value input button | Convert to setting value inputscreen. | |
No. | Contents | Description | Additional |
j | Operation Screen | Transfer operation screen. | |
k | Set Function | Transfer function and FIX settingscreen. | |
l | Set Reserve | Transfer reserve setting screen. | |
m | Display Graph | Transfer graph setting screen. | |
n | Set Pattern | Transfer pattern setting screen. | |
o | Set Display | Transfer display setting screen. |
No. | Contents | Description | Additional |
① | Main Button | ||
② | Pattern No. | Display running pattern number. | ▶ Push button and set pattern number for operating. |
③ | Segment | Display programmed segment count of the present pattern. | ▶ Program pattern input |
④ | Time Button | Display present time. Time button is pushed, turn off LCD screen. Even if turn off the light automatically, controller still working and turn on the light if user touch any point. | ▶ Automatic turn off setting time ▶ Backlight time is setting 10 minutes. ▶ Present time is setting |
⑤ | RUN Button | Display confirmable box for program running. | ▶ See (Figure1-4). |
⑥ | Present Humi. | Display present humidity. | |
⑦ | PresentTemp. | Display present temperature. | |
⑧ | Next Button | Transfer to Figure 1-15 |
No. | Contents | Description | Additional |
① | Confirmable Box | Program run or not. | ▶ Clicked YES button : Program Start ▶ Clicked NO button : Transfer at Figure 1-3. |
② | Pattern End | Blinking at the end of setting program. | ▶ if user touch any point, PT END disappear. ▶ If Pattern End message is disappeared,Relay status transfer into OFF. |
③ | Start Run Time | Display start run time at setting reservation. | ▶ The reserved time must be later than present time. |
④ | Reservation | Blinking at setting reservation. | |
⑤ | RUN Button | Reserve status,Press this button to run at once. | ▶ Press this button,Reservation will be canceled,and transfer at running status. |
No. | Contents | Description | Additional |
① | Pattern Name | Display pattern name of the present running. | |
② | Next Button | Transfer at Figure 1-8. | |
③ | Temp. Status Display Lamp | Display present Temp. status. | |
④ | Humi. Status Display Lamp | Display present Humi. status. | |
⑤ | Temp. SP Value | Display present Temp. SP(Set Point). | |
⑥ | Humi. SP Value | Display present Humi. SP(Set Point). | |
⑦ | Pattern/Segment | Display pattern and segment number. | |
⑧ | Hold Button | Hold on or off. | ▶ Hold button’s color transferred to Blue at HOLD ON. |
⑨ | Step Button | Present segment stop running, Move next segment. | |
⑩ | Stop Button | Display confirmable box. | |
11 | Running Time | Display total running time. |
No. | Contents | Description | Additional |
① | Temp. MVOUT | Display Temp. MVOUT. | |
② | Humi. MVOUT | Display Humi. MVOUT. | |
③ | Running Pattern | Display program pattern and segment number. | |
④ | Pattern Repeat | Display pattern repeat status. repeated count / total repeat count | |
⑤ | Segment Repeat | Display segment repeat status. repeated count / total repeat count | |
⑥ | Remain Time | Display remain time of the present running segment. | |
⑦ | Status Display Lamp | ON status : Red. OFF status : Gray. | |
⑧ | Next Button | Transfer at Figure 1-15. | TEMI850 Tranfer at Figure1-7. |
No. | Contents | Description | Additional |
① | Temp. SP Value | Display present Temp. SP (Set Point). | ▶ See 1.2 Input Setting Value Input SP(Set Point). |
② | Humi. SP Value | Display present Humi. SP (Set Point). | |
③ | Time | Display present Time,Same as program. | |
④ | Fix Run Stop | Blinking at Fix Stop. | ▶ if user touch any point, FIX END disappear. |
⑤ | Reservation | Blinking at setting reservation. | |
⑥ | RUN Button | Reserve status,Press this button to run at once. | ▶ Press this button,Reservation will be canceled,and transfer at running status. |
No. | Contents | Description | Additional |
j | T-AT | Temp. side : Auto Tuning On/Off. | ▶ T-AT button’s color transferred to blue at Auto Tuning of the Temp. side. |
k | H-AT | Humi. side : Auto Tuning On/Off. | ▶ H-AT button’s color transferred to blue at Auto Tuning of the Humi. side. |
l | PID Group No. | Display PID Group No. | |
m | Next Button | Transfer at Figure 1-15. | TEMI850 Tranfer at 1-13 |
No. | Contents | Description | Additional |
j | NEXT Button | If Prog Running,Transfer to Figure 1-7 Program Run – First Screen; If Fix Running, Transfer to Figure 1-13 Fix Run – First Screen. | ▶ See Figure 1-7 Program Run – First Screen ▶ See Figure 1-13 Fix Run – First Screen |
k | PAR. Button | Tranfer to Monitor Graph Attribute Setting Screen. | |
l | DEL Button | Delete Monitor Graph。 | Deleted Graph won’t be resumed. |
m | TEMP Graph Display or not Button. | To select Display or Hide TEMP PV and SP。 | ▶ TEMP PV: In Red。 ▶ TEMP SP: In Green。 |
n | HUMI Graph Display or not Button. | To select Display or Hide HUMI PV and SP。 | ▶ HUMI PV: In blue。 ▶ HUMI SP: In black。 |
o | LEFT Move Button | Left move window,to see earlier Graph. | |
p | RIGHT Move Button | Right move window,to see later Graph. |
No. | Contents | Description | Additional |
j | EXIT Button | Back to Monitor Graph Screen. | |
k | Sample period. | Select a sample period for Monitor Graph.Because of the limited space of memorizer of TEMI880.different Sample period will bring on different record time. | ▶ Limited Record Time: 1s:6 hours. 10s:60 hours. 30s:180 hours. 60s:360 hours. |
l | FIN. MODE | STOP:Space used up,stopping record at once. REP.:(Repeat) When space used up, record operation will continue,but the old record will be coverd | |
m | TEMP RANGE | Set a Y coordinate for Temp,Range High,and Range Low | ▶ Different range width will bring a different outline effect of the Graph. |
n | HUMI RANGE | Set a Y coordinate for Humi,Range High,and Range Low |
No. | Contents | Description | Additional | |||
j | Operation Mode | Select Program mode or Fix mode. | ▶ PROG : 1.5.1 Program Stop Screen ▶ FIX :1.5.3 Fix Stop Screen ☞ It isn’t changed at the running. | |||
k | Power Mode | Setup when re-start after power off. ▶ Recognize to power off if it is more than 3 seconds. ▶ If re-start before 3 seconds, It will run to Hot mode Automatically. | ▶ The running after power on | |||
Before Power Off | PROG/FIX STOP | PROG RUN | FIX RUN | |||
STOP | PROG/FIX STOP | PROG STOP | FIX STOP | |||
COLD | PROG/FIX STOP | Run from first segment. | FIX RUN | |||
HOT | PROG/FIX STOP | Run from running segment time before power off. | FIX RUN | |||
l | Key Lock | Key Lock/Unlock. | ||||
m | LightTime | Set time of the backlight. | ▶ The default backlight time is setting 10 minutes. | |||
No. | Contents | Description | Additional |
j | Operation Hour(H) | Set Hour. | |
k | Operation Minute(M) | Set Minute. | |
l | Time Mode | Set Time Mode On/Off. | ▶ OFF : Fix running is stopped at the STOP button clicked. ▶ ON : It is stop after running time (set j,k) automatically. |
m | Temp. Slope | Temp. up/down slope. | ▶ Temp./Humi. value goes to the SP gradually. It is applied both of up and down. |
n | Humi. Slope | Humi. up/down slope. |
No. | Contents | Description | Additional |
j | Current Time | Set and display present date & time. | |
k | Reserve Time | Set and display reservation date & time. | |
l | Reserve Mode | Select Reservation On or Off. | ▶ See Figure 1-6, Figure 1-12 |
No. | Contents | Description | Additional | |
j | TIME Button | Change time scale. | ▶ Push time button => change time scale. | |
0H 1H 2H 3H 4H | ||||
↓ | ||||
0H 3H 6H 9H 12H | ||||
↓ | ||||
0H 6H 12H 18H 24H | ||||
↓ | ||||
0H 12H 24H 36H 48H | ||||
↓ | ||||
0H 24H 48H 72H 96H | ||||
k | MAIN | Transfer to Main Screen | ||
l | Pattern No. | Set pattern number. | ||
m | TEMP SP | Display/Hide Temp SP Graph。 | ▶Temp. SP : Display red. | |
n | HUMI SP | Display/Hide Humi SP Graph。 | ▶ Humi. SP : Display blue. | |
o | Previous/Next Button | Move previous/next page. | ▶ See1.5.5.1 Monitor Graph Screen . | |
No. | Contents | Description | Additional |
j | PROG Pattern EDIT | Move toedit pattern screen. | |
k | Set Repeat | Move torepeat setting screen. | |
l | Edit File | Move to edit file screen. | |
m | Time Signal | Move totime signal setting screen. | ▶See 1.9.4 Time Signal Seeting Screen |
n | Set Wait | Move towait operation setting screen. | |
o | Pattern Name | Move topattern name setting screen. | |
⑦ | Message | Move to Prog File Message Screen. |
No. | Contents | Description | Additional | |
j | Segment No. | A case where insert or delete segment, Clicked this button. | ▶ Segment color is displayed green at the buttonclicked. In this case, you can segmentinsert or delete. | |
k | Segment Temp. | Set segment temperature. | ||
l | Segment Humi. | Set segment humidity. | ||
m | Segment Time | Set segment time. | ▶ Input method of the time. | |
Set Time | Display | |||
OFF | -0.01 | |||
1 minute | 0.01 | |||
30 minutes | 0.30 | |||
1 hour | 1.00 | |||
n | PG UP | Move previous page. | ||
o | PG DN | Move next page. | ||
p | Insert Segment | Insert segment. | ||
q | Delete Segment | Delete segment. | ||
r | Time Signal | Set time signal. | ||
s | Pattern No. | Set pattern number. | ☞ Change pattern number of | |
No. | Contents | Description | Additional |
j | Pattern No. | Set pattern number. | Change pattern number of |
k | Pattern Repeat No. | Set pattern repeat number. | “0” is repeating unlimitedly. |
l | Link Pattern | When the finish the runningpattern, set the next patternnumber continually. | “0” is no Link pattern. |
m | Start Segment parts. | Set the start segment numberof repeat | ▶ Notice: Start Segment ≤ End Segment, Otherwise,system will buzz “bi bi” to alarm. |
n | End Segment | Set the end segment numberof repeat parts. | |
o | Count Segment | Set the number of repeat parts. |
Set value of the pattern repeat | Process order |
▶ Repeat 3 times Pattern 1,then jump to Pattern 2. | |
▶ Repeat 5 times Pattern 3,then jump to Pattern 4. | |
▶ Repeat 5 times Pattern 3,then stop. | |
▶ Repeat Pattern 3 forever. PT CYC = 0,Indicate repeat this pattern forever,even if a valid LINK PT NO is setted | |
▶ Repeat Pattern 1 forever. Link to the pattern itself,though PT CYC =3,it is inessential. |
No. | Set value of the repeat parts | Process order |
Example 1 | ① 01→02→03→04 ② 02→03→04 ③ 03→04→05 ④ 03→04→05 ⑤ 01→02→03→04 ⑥ 01→02→03→04 ⑦ 04→05→06 ⑧ 04→05→06→07→08 | |
Example 2 | ① 01→02→03→04 ② 02→03→04 ③ 04→05→06 ④ 04→05→06→07→08 | |
Example 3 | ① 01→02→03→04→05→06→07→08 ② 07→08 ③ 04→05→06 ④04→05→06→07→08 | |
Example 4 | ① 01→02→03→04→05 ②03→04→05→06→07→08 | |
Example 5 | ① 01→02→03→04→05→06→07 ②03→04→05 ③03→04→05→06→07→08 | |
Example 6 | ① 01→02→03→04→05→06→07 ② 05 ③ 05→06→07→08 | |
Example 7 | ① 01→02→03→04→05→06→07→08 |
No. | Contents | Description | Additional |
j | Source Pattern | Set source pattern number. | |
k | Destination Pattern | Set destination pattern number. | |
l | Delete Pattern | Set pattern number for elimination. | |
m | Delete Result | Delete Result indication. | |
n | Execute Result | Display execute result. | |
o | Pattern Copy Button | Pattern copy. | ▶ Copy Source Pattern to Destination Pattern. ▶ You can not copy if source pattern is empty. |
p | Pattern Delete Execute Button | One Pattern Delete operation. | Deleted Pattern can’t be resumed. |
q | All Pattern Delete Execute Button | All Pattern Delete operation. | Deleted all Pattern can’t be resumed |
No. | Contents | Description | Additional |
① | On Time | Set wait time for the time signaloutput. | → No use time signal. ▶ NO : 01 → Use time signal. ▶ NO : 02 ~ 07 → Use time signal from On time to Off time. |
② | Off Time | Set running time for the time signal output. |
Time Signal Operation | |||
▶ Set value of the program pattern | ▶ Set value of the time signal | ||
▶ Time signal operation within segment | |||
Setting | Time signal operation | ||
N SEG TIME >= ON TIME + OFF TIME ☞ It has not influence on next segment. | 1.ON TIME =00.00 OFF TIME = 04.00 (Time SignalNo:04 ) | ||
2. ON 时间 = 02.00 OFF 时间 = 02.00 (Time Signal:05) | |||
N SEG TIME < ON TIME + OFF TIME | 2.ON 时间=00.00 OFF 时间 = 10.00 (Time Signal:06) | ||
3.ON 时间 = 04.00 OFF 时间 = 06.00 (Time Signal:07) | |||
No. | Contents | Description | Additional |
j | Wait Mode | Set On/Off for using Wait. | ▶ “OR” Condition → Temp. or Humi. PV does not reachthe wait area before Wait Time. ▶ “AND” Condition → Temp. and HUmi. PV reach the wait area before Wait Time. |
k | Temp. Zone | Set a Temp. zone. | |
l | Humi. Zone | Set a Humi. zone. | |
m | Wait Time | Set a Wait time. | ▶ Initial time (00.00) → Wait to infinity. |
The relation of the wait operation and wait time ▶ WAIT ZONE : Temperature → TEMP ZONE, Humidity → HUMI ZONE. |
① A case where wait release within Wait Time |
②A case where PV does not enter WZ(Wait Zone) within Wait Time |
No. | Contents | Description | Additional |
j | Pattern Name Button | Set pattern name. | ▶ Max : use 10 character. |
k | Page Move Button | Move pattern page. | ▶ 1 Page : Display 5 pattern name. |
No. | Contents | Description | Additional |
j | MAIN Button | Move to Main Screen. | |
k | Tuning Key | Display or hide AT button. | ▶ Display or hide T-AT/H-AT button of the Figure 1-13. Figure1-14 |
l | Refresh Frequency | Adjust the LCD’s Refresh Frequyncy | ▶ Factory used to debug only,not recommend user or the third party to adjust it. |