TEMI880

- Color type

TEMI850
- Mono type
操
INSTRUCTION MANUAL
TEMPERATURE • HUMIDITY
PROGRAMMABLE CONTROLLER
Contents


1.User Manual
1.1 Setup Button
▶ TEMI880 is easy to install & modify by touch screen as well as the map type menu. User enablesto grasp the operation of control system with easy through graph on vivid LCD screen.
1.1.1Initial setup button
▶ Table 1-1. General Setting Button

No.
Button
Function
1
Main
Change the page.
2
General
execute button
It is used general execution orselection.
3
Page Up/Down button
It is used page conversion on thesame screen.
4
Next page
button
It is used to convert to next page.
5
Setting value
select button
It is used setting value selection.
6
Setting value
input button
Convert to setting value inputscreen.

1.1.2Input setting value
▶ If user push setting value input button in the Table 1-1(General Setting Button), User can input
data value.
▶ If user input data which is invalid, it makes 3 times buzzer, also there is “-LIMIT ERR” messageon the setting value screen.
① Input key for the Digit setting②Input key for the Pattern & DI Error name setting
Touch Key Unlock
▶ If KEY LOCK is ON, Not input setting value. Therefore, Input setting value after KEY LOCKconvert OFF.
▶ For more information, See 1.6 Operation Setting Screen.
1.1.3Input error
▶ User can check a number of buzzer times.
1 time : Succeed set the data.
3 times : Failed set the data or the operation order is impropriety,or the key is locked .
Notice Items.
▶ Using the touch screen. Please do not use any pen/pencil, nail or any other sharp material.Otherwise, the touch screen might be broken or out of order.
▶ In order to avoid invalid input. When you set the parameter,Please according to the order High ,Middle, Low, where it is needed.
1.2 Input Setting Value
▶ All of the Input value is setting by use setting value, pattern name and time signal input key.
▶ If user push setting value input button in the Table 1-1(General Setting Button), User can input
data value.
▶ For the Pattern Name input key, See 1.9.6 Pattern Name Setting Screen.
1.2.1Function of setting value input key
① ⑨⑧
③
② ④
⑦


⑥⑤
Figure 1-1. Function of setting value input key
① Parameter.
② Valid data range.
③ “ERROR!”will be displayed when user input invalid setting value.
④ Cancel input setting value.
⑤ Clear input data.
⑥ Decimal point.
⑦ Save input data and go to main page.
⑧ Modify the input data, delete one by one.
⑨ Input sign(+ or -).


1.3 Block diagram of the basic processing

Power On


【 PROG STOP】 【Initial Screen】
【FIX STOP】 【Main Screen】
【DI TROUBLE SCREEN】【PROGRAM SET】


1.4 Main Screen
▶ This screen is center of the total screen transfer.
Figure 1-2. Main Screen
jm
kn
lo

No.
Contents
Description
Additional
j
Operation Screen
Transfer operation screen.
k
Set Function
Transfer function and FIX settingscreen.
l
Set Reserve
Transfer reserve setting screen.
m
Display Graph
Transfer graph setting screen.
n
Set Pattern
Transfer pattern setting screen.
o
Set Display
Transfer display setting screen.

1.5 Operation Screen
▶At the running TEMI880, Display status and information.
1.5.1Program Stop Screen
Figure 1-3. PROG STOP Screen
①⑧
②⑦
③⑥


④⑤

No.
Contents
Description
Additional
①
Main Button
Transfer to 1.4 Main Screen.
②
Pattern No.
Display running pattern number.
▶ Push button and set pattern number
for operating.
③
Segment
Display programmed segment
count of the present pattern.
▶ Program pattern input
④
Time Button
Display present time.
Time button is pushed, turn off
LCD screen. Even if turn off the
light automatically, controller
still working and turn on the light
if user touch any point.
▶ Automatic turn off setting time
▶ Backlight time is setting 10 minutes.
▶ Present time is setting
⑤
RUN Button
Display confirmable box for
program running.
▶ See (Figure1-4).
⑥
Present Humi.
Display present humidity.
⑦
PresentTemp.
Display present temperature.
⑧
Next Button
Transfer to Figure 1-15

Confirmable Box
At this CONFIRM window, you can confirm if you really wantdoing job or not.
예) PROG(or FIX) STOP ↔ PROG(or FIX) RUN
HOLD, STEP, TUNING OFF ↔ HOLD, STEP, TUNING ON
pattern copy, pattern delete etc.
Figure 1-4. PROG STOP – ConfirmFigure 1-5. PROG STOP – Pattern End
①②
Figure 1-6. PROG STOP – Reserved
③


④⑤

No.
Contents
Description
Additional
①
Confirmable Box
Program run or not.
▶ Clicked YES button : Program Start
▶ Clicked NO button : Transfer
at Figure 1-3.
②
Pattern End
Blinking at the end of setting
program.
▶ if user touch any point, PT END
disappear.
▶ If Pattern End message is disappeared,Relay status transfer into OFF.
③
Start Run Time
Display start run time at setting
reservation.
▶ The reserved time must be later than present time.
④
Reservation
Blinking at setting reservation.
⑤
RUN Button
Reserve status,Press this button to run at once.
▶ Press this button,Reservation will be canceled,and transfer at running status.


1.5.2Program Run Screen
①
Figure 1-7. Program Run – First Screen
②
③
⑤
④
⑥
⑦
11
⑧⑨⑩

No.
Contents
Description
Additional
①
Pattern Name
Display pattern name of the present
running.
②
Next Button
Transfer at Figure 1-8.
③
Temp. Status
Display Lamp
Display present Temp. status.
④
Humi. Status
Display Lamp
Display present Humi. status.
⑤
Temp. SP Value
Display present Temp. SP(Set Point).
⑥
Humi. SP Value
Display present Humi. SP(Set Point).
⑦
Pattern/Segment
Display pattern and segment number.
⑧
Hold Button
Hold on or off.
▶ Hold button’s color transferred to
Blue at HOLD ON.
⑨
Step Button
Present segment stop running, Move
next segment.
⑩
Stop Button
Display confirmable box.
11
Running Time
Display total running time.


Figure 1-8. Program Run – Second Screen
⑧
①②
③
④⑦
⑤
⑥

No.
Contents
Description
Additional
①
Temp. MVOUT
Display Temp. MVOUT.
②
Humi. MVOUT
Display Humi. MVOUT.
③
Running Pattern
Display program pattern and
segment number.
④
Pattern Repeat
Display pattern repeat status.
repeated count / total repeat count
⑤
Segment Repeat
Display segment repeat status.
repeated count / total repeat count
⑥
Remain Time
Display remain time of the present
running segment.
⑦
Status Display
Lamp
ON status : Red.
OFF status : Gray.
⑧
Next Button
Transfer at Figure 1-15.
TEMI850 Tranfer at Figure1-7.


1.5.3Fix Stop Screen
Figure 1-9. Fix Stop ScreenFigure 1-10. Fix Stop – Confirm


① ② ③
Figure 1-11. Fix Stop – Time EndFigure 1-12. Fix Stop – Reserved
④⑤⑥

No.
Contents
Description
Additional
①
Temp. SP
Value
Display present Temp. SP (Set Point).
Input SP(Set Point).
②
Humi. SP
Value
Display present Humi. SP (Set Point).
③
Time
Display present Time,Same as program.
④
Fix Run Stop
Blinking at Fix Stop.
▶ if user touch any point, FIX END
disappear.
⑤
Reservation
Blinking at setting reservation.
⑥
RUN Button
Reserve status,Press this button to run at once.
▶ Press this button,Reservation will be canceled,and transfer at running status.


1.5.4Fix Run Screen
m
Figure 1-13. Fix Run – First ScreenFigure 1-14. Fix Run – Second Screen


jkl

No.
Contents
Description
Additional
j
T-AT
Temp. side : Auto Tuning On/Off.
▶ T-AT button’s color transferred to
blue at Auto Tuning of the Temp. side.
k
H-AT
Humi. side : Auto Tuning On/Off.
▶ H-AT button’s color transferred to
blue at Auto Tuning of the Humi. side.
l
PID Group No.
Display PID Group No.
m
Next Button
Transfer at Figure 1-15.
TEMI850 Tranfer at 1-13


1.5.5Monitor graph & attribute .
1.5.5.1 Monitor Graph Screen
▶ This function is only for TEMI880.
Figure 1-15 Monitor Graph Screen
j
k
l
m
n
o
p

No.
Contents
Description
Additional
j
NEXT Button
If Prog Running,Transfer to Figure 1-7 Program Run – First Screen;
If Fix Running, Transfer to Figure 1-13 Fix Run – First Screen.
▶ See Figure 1-7 Program Run – First Screen
▶ See Figure 1-13 Fix Run – First Screen
k
PAR. Button
Tranfer to Monitor Graph Attribute Setting Screen.
l
DEL Button
Delete Monitor Graph。
Deleted Graph won’t be resumed.
m
TEMP Graph Display or not Button.
To select Display or Hide TEMP PV and SP。
▶ TEMP PV: In Red。
▶ TEMP SP: In Green。
n
HUMI Graph Display or not Button.
To select Display or Hide HUMI PV and SP。
▶ HUMI PV: In blue。
▶ HUMI SP: In black。
o
LEFT Move Button
Left move window,to see earlier Graph.
p
RIGHT Move Button
Right move window,to see later Graph.


1.5.5.2 Monitor graph attribute setting.
Figure 1-16 Monitor Graph Attribute Setting Screen
①
②
③
④
⑤

No.
Contents
Description
Additional
j
EXIT Button
Back to Monitor Graph Screen.
k
Sample period.
Select a sample period for Monitor Graph.Because of the limited space of memorizer of TEMI880.different Sample period will bring on different record time.
▶ Limited Record Time:
1s:6 hours.
10s:60 hours.
30s:180 hours.
60s:360 hours.
l
FIN. MODE
STOP:Space used up,stopping record at once.
REP.:(Repeat) When space used up, record operation will continue,but the old record will be coverd
m
TEMP RANGE
Set a Y coordinate for Temp,Range High,and Range Low
▶ Different range width will bring a different outline effect of the Graph.
n
HUMI RANGE
Set a Y coordinate for Humi,Range High,and Range Low


1.6 Operation Setting Screen
▶ This is the screen about additional function of the general machine and additional setting at Fix Run.
Figure 1-17. Operation Setting – First Screen
①
②
③
m

No.
Contents
Description
Additional
j
Operation
Mode
Select Program mode or Fix
mode.
☞ It isn’t changed at the running.
k
Power Mode
Setup when re-start
after power off.
▶ Recognize to power off if it is
more than 3 seconds.
▶ If re-start before 3 seconds,
It will run to Hot mode
Automatically.
▶ The running after power on
Before
Power Off
PROG/FIX
STOP
PROG RUN
FIX
RUN
STOP
PROG/FIX
STOP
PROG STOP
FIX
STOP
COLD
PROG/FIX
STOP
Run from first
segment.
FIX
RUN
HOT
PROG/FIX
STOP
Run from running
segment time
before power off.
FIX
RUN
l
Key Lock
Key Lock/Unlock.
m
LightTime
Set time of the backlight.
▶ The default backlight time is setting 10 minutes.


Figure 1-18. Operation Setting – Second Screen
①
②
③
④
⑤

No.
Contents
Description
Additional
j
Operation
Hour(H)
Set Hour.
k
Operation
Minute(M)
Set Minute.
l
Time Mode
Set Time Mode On/Off.
▶ OFF : Fix running is stopped at the
STOP button clicked.
▶ ON : It is stop after running time
(set j,k) automatically.
m
Temp. Slope
Temp. up/down slope.
▶ Temp./Humi. value goes to the SP gradually.
It is applied both of up and down.
n
Humi. Slope
Humi. up/down slope.


1.7 Reserve Setting Screen
▶ This is the screen about time and reserved operation setting.
Figure 1-19. Reserve Setting Screen
①
②
③

No.
Contents
Description
Additional
j
Current Time
Set and display present date & time.
k
Reserve Time
Set and display reservation date & time.
l
Reserve Mode
Select Reservation On or Off.
▶ See Figure 1-6, Figure 1-12


1.8 Graph Setting Screen
▶ This is screen display graph for the input pattern at 1.9.1 Edit Pattern Screen.
Figure 1-20. Graph Setting Screen
①
②
③
④
⑤


⑥

No.
Contents
Description
Additional
j
TIME Button
Change time scale.
▶ Push time button
=> change time scale.
0H 1H 2H 3H 4H
↓
0H 3H 6H 9H 12H
↓
0H 6H 12H 18H 24H
↓
0H 12H 24H 36H 48H
↓
0H 24H 48H 72H 96H
k
MAIN
Transfer to Main Screen
l
Pattern No.
Set pattern number.
m
TEMP SP
Display/Hide Temp SP Graph。
▶Temp. SP : Display red.
n
HUMI SP
Display/Hide Humi SP Graph。
▶ Humi. SP : Display blue.
o
Previous/Next
Button
Move previous/next page.


1.9 Pattern Setting Screen
▶ This is the screen about pattern setting.
Figure 1-21. Pattern Setting Screen
⑦
jm
kn
lo

No.
Contents
Description
Additional
j
PROG Pattern EDIT
Move toedit pattern screen.
k
Set Repeat
Move torepeat setting screen.
l
Edit File
Move to edit file screen.
m
Time Signal
Move totime signal setting screen.
▶See 1.9.4 Time Signal Seeting Screen
n
Set Wait
Move towait operation setting screen.
o
Pattern
Name
Move topattern name setting screen.
⑦
Message
Move to Prog File Message Screen.

1.9.1Edit Pattern Screen
Figure 1-22. Edit Pattern Screen
s


jr
k
l
m


nopq

No.
Contents
Description
Additional
j
Segment No.
A case where insert or delete
segment, Clicked this button.
▶ Segment color is displayed green at the buttonclicked. In this case, you can segmentinsert or delete.
k
Segment Temp.
Set segment temperature.
l
Segment Humi.
Set segment humidity.
m
Segment Time
Set segment time.
▶ Input method of the time.
Set Time
Display
OFF
-0.01
1 minute
0.01
30 minutes
0.30
1 hour
1.00
n
PG UP
Move previous page.
o
PG DN
Move next page.
p
Insert Segment
Insert segment.
q
Delete Segment
Delete segment.
r
Time Signal
Set time signal.
s
Pattern No.
Set pattern number.
☞ Change pattern number of

you are not able to edit a prog pattern when it is running.If you attempt to do this,
TEMI880 will buzz “bi bi”to alarm.


1.9.1.1 Insert Method of the Segment
▶ Example => Insert segment.
Figure 1-23. Before Insert SegmentFigure 1-24. After Insert Segment
1) Clicked “2”areas of Figure 1-23, then the background of “2” character color is changed green to indication editable status.
2) Clicked insert button(p ) of Figure 1-22. Edit Pattern Screen, a segment of the same value is
Inserted,behind Segment 2.
1.9.1.2 Delete Method of the Segment
▶ Example => Delete segment.
Figure 1-25. Before Delete SegmentFigure 1-26. After Delete Segment
1) Clicked “2”areas of Figure 1-25, then the background of “2” character color is changed green to indication editable status.
2) Clicked delete button(q) of Figure 1-22. Edit Pattern Screen, “2” segment is deleted.
A case where segment do not insert or delete.
When a prog pattern is running.


1.9.1.3 Move Screen
obutton of Figure 1-22. Edit Pattern Screen


nbutton of Figure 1-22. Edit Pattern Screen
1.9.1.4 Set Method of the Time Signal
▶ This is the screen about time signal setting.
Figure 1-27. Before Time Signal Setting
①


As Figure 1-27 above,Clickeddot-rectangle areas of(①),will display Figure
1-28. Time Signal Setting Key.
Figure1-28. Time Signal Setting Key.
1) SEGx: Indicate the segment No.of a Prog(1 ~ 100)
2) TSx: Indicate one of TS(TS1~ TS4)of this segment of the prog.
3) Input a number , the range is form 0 to 7.(0 ~ 7).then press ENT.
A kind of the input time signal : See 1.9.4 Time Signal Setting Screen.


1.9.1.5 Prog File Message Screen
▶ Display number of the used pattern and segment.
Figure 1-29. File Information Screen
▶ How many Progs and Segments have been used will be displayed in this screen.
▶ Press“PRO.SET”to move to Figure 1-21 Pattern Setting Screen.
1.9.2Repeat Setting Screen
Figure 1-30. Repeat Setting Screen
j
kl
m
n
o

No.
Contents
Description
Additional
j
Pattern No.
Set pattern number.
Change pattern number of
k
Pattern Repeat
No.
Set pattern repeat number.
“0” is repeating unlimitedly.
l
Link Pattern
When the finish the runningpattern, set the next patternnumber continually.
“0” is no Link pattern.
m
Start Segment
parts.
Set the start segment numberof repeat
▶ Notice:
Start Segment ≤ End Segment, Otherwise,system will buzz “bi bi” to alarm.
n
End Segment
Set the end segment numberof repeat parts.
o
Count Segment
Set the number of repeat parts.

1.9.2.1Set Pattern Repeat

Set value of the pattern repeat
Process order
▶ Repeat 3 times Pattern 1,then jump to Pattern 2.
▶ Repeat 5 times Pattern 3,then jump to Pattern 4.
▶ Repeat 5 times Pattern 3,then stop.
▶ Repeat Pattern 3 forever.
PT CYC = 0,Indicate repeat this pattern forever,even if a valid LINK PT NO is setted
▶ Repeat Pattern 1 forever.
Link to the pattern itself,though PT CYC =3,it is inessential.

1.9.2.2 Process Order about Setting Method of the Repeat Parts(Segment Repeat)
▶ Examples,
▶ It is assumed a pattern has already setted its 1 segment ~ 8 segment(01 → 02 → 03 → 04 → 05 → 06 → 07 →08)

No.
Set value of the repeat parts
Process order
Example
1
① 01→02→03→04
② 02→03→04
③ 03→04→05
④ 03→04→05
⑤ 01→02→03→04
⑥ 01→02→03→04
⑦ 04→05→06
⑧ 04→05→06→07→08
Example
2
① 01→02→03→04
② 02→03→04
③ 04→05→06
④ 04→05→06→07→08
Example 3
① 01→02→03→04→05→06→07→08
② 07→08
③ 04→05→06
④04→05→06→07→08
Example 4
① 01→02→03→04→05
②03→04→05→06→07→08
Example 5
① 01→02→03→04→05→06→07
②03→04→05
③03→04→05→06→07→08
Example 6
① 01→02→03→04→05→06→07
② 05
③ 05→06→07→08
Example 7
① 01→02→03→04→05→06→07→08


1.9.3Edit File Screen
Figure 1-31. Edit File Screen
m
jn
ko


lp
q


No.
Contents
Description
Additional
j
Source Pattern
Set source pattern number.
k
Destination
Pattern
Set destination pattern number.
l
Delete Pattern
Set pattern number for elimination.
m
Delete Result
Delete Result indication.
n
Execute Result
Display execute result.
o
Pattern Copy
Button
Pattern copy.
▶ Copy Source Pattern to Destination Pattern.
▶ You can not copy if source pattern is
empty.
p
Pattern Delete Execute Button
One Pattern Delete operation.
Deleted Pattern can’t be resumed.
q
All Pattern Delete Execute Button
All Pattern Delete operation.
Deleted all Pattern can’t be resumed


1.9.4Time Signal Setting Screen
Figure 1-32. TS Setting – First ScreenFigure 1-33. TS Setting – Second Screen
① ②

No.
Contents
Description
Additional
①
On Time
Set wait time for the time signaloutput.
→ No use time signal.
▶ NO : 01
→ Use time signal.
▶ NO : 02 ~ 07
→ Use time signal from On time to
Off time.
②
Off Time
Set running time for the time
signal output.

Time Signal Operation
▶ Set value of the program pattern
▶ Set value of the time signal
▶ Time signal operation within segment
Setting
Time signal operation
N SEG TIME
>= ON TIME
+ OFF TIME
☞ It has not
influence on
next segment.
1.ON TIME =00.00
OFF TIME = 04.00
(Time SignalNo:04 )
2. ON 时间 = 02.00
OFF 时间 = 02.00
(Time Signal:05)
N SEG TIME
< ON TIME
+ OFF TIME
2.ON 时间=00.00
OFF 时间 = 10.00
(Time Signal:06)
3.ON 时间 = 04.00
OFF 时间 = 06.00
(Time Signal:07)

1.9.5Wait Operation Setting Screen
Figure 1-34. Wait Operation Setting Screen
j
k
l
m

No.
Contents
Description
Additional
j
Wait Mode
Set On/Off for using Wait.
▶ “OR” Condition
→ Temp. or Humi. PV does not reachthe wait area before Wait Time.
▶ “AND” Condition
→ Temp. and HUmi. PV reach the wait
area before Wait Time.
k
Temp. Zone
Set a Temp. zone.
l
Humi. Zone
Set a Humi. zone.
m
Wait Time
Set a Wait time.
▶ Initial time (00.00)
→ Wait to infinity.

The relation of the wait operation and wait time
▶ WAIT ZONE : Temperature → TEMP ZONE, Humidity → HUMI ZONE.
① A case where wait release within Wait Time
②A case where PV does not enter WZ(Wait Zone) within Wait Time

1.9.6Pattern Name Setting Screen
Figure 1-35. Pattern Name Setting Screen
k
j


No.
Contents
Description
Additional
j
Pattern Name
Button
Set pattern name.
▶ Max : use 10 character.
k
Page Move
Button
Move pattern page.
▶ 1 Page : Display 5 pattern name.

1.9.6.1 The Method of the Pattern Name Setting
▶ The example of the pattern 02name setting : “PROG2”.
Figure 1-36. Before Pattern Name Setting
⑥⑤
Figure 1-37. Pattern Name Setting Key
①


② ③④⑦
▶ If user touch dot-rectangle areas(①) of Figure 1-36. Before Pattern Name Setting, Display Figure1-37. Pattern Name Setting Key.
▶ The order of the pattern name setting :③→③→⑤→②→⑤→⑤→④→⑥→⑤→⑤→⑤→⑦。
1.10 Display Setting Screen
▶ This is the screen of the display setting.
Figure 1-38. Display Setting Screen
①
②
③

No.
Contents
Description
Additional
j
MAIN Button
Move to Main Screen.
k
Tuning Key
Display or hide AT button.
▶ Display or hide T-AT/H-AT button of the
Figure 1-13. Figure1-14
l
Refresh Frequency
Adjust the LCD’s Refresh Frequyncy
▶ Factory used to debug only,not recommend user or the third party to adjust it.