Proper use and maintenance is the key to the normal operation of the machine, to extend the service life, to maintain high performance and high indicators. Maintain and maintain the ICP spectrometer as per the following requirements and methods in daily analysis.
您在使用ICP-OES光谱仪和保养时还应注意以下几点:
You should also pay attention to the following points when using ICP-OES spectrometer and maintenance:
1、退出OIA软件并关闭激发台光源开关(红色按钮)
1, exit the OIA software and turn off the ignition light switch (red button)
2、取下燃烧室两侧的螺丝,向上移开燃烧室。
2, remove the screws on both sides of the combustion chamber, the combustion chamber.
3、用专用毛刷(或者脱脂棉)清扫钨电极周围和燃烧室里面的黑色废渣(严禁使用酒精等液体)。
3, with a special brush (or absorbent cotton) to clean the tungsten electrode and the combustion chamber inside the black waste residue (prohibit the use of alcohol and other liquids).
4、清理完毕装上燃烧室,打开光谱仪软件和激发光源(红)后,氩气吹扫十几分钟。
4, clean up the combustion chamber installed, open the spectrometer software and excitation light source (red), argon blowing more than ten minutes.
注意:不要将大块物体落入排气孔内,以免堵塞排气管道。
Note: do not put large objects into the exhaust hole, so as not to block the exhaust pipe.
只要做到以上几点,然后熟悉熟练的应用,你就会发现原来天瑞ICP光谱仪轻松维护是如此的简单。
As long as the above points, and then familiar with skilled application, you will find that the original ICP spectrometer easy maintenance is so simple.