企业档案
会员类型:旗舰版
已获得阿仪网信誉 等级评定
(0 -40)基础信誉积累,可浏览访问
(41-90)良好信誉积累,可接洽商谈
(91+ )优质信誉积累,可持续信赖
阿仪网旗舰版会员:9年
最后认证时间:
注册号: 【已认证】
法人代表: 【已认证】
参观次数:712556
手机网站:http://wap.app17.com/c134335/
旗舰版网址:http://guorui.app17.com
技术文章
订制英文版空气微生物采样器
点击次数:2742 发布时间:2024/8/28 16:14:58
GR-1353型空气微生物采样器
1 Product Overview

青岛国瑞英文版空气微生物采样器
The GR-1353 Air Microbial Sampler (hereinafter referred to as the sampler) is a sampling instrument specially developed by our company for collecting microorganisms in the air. It is based on the principle of Anderson's impact method, extracting air at a certain flow rate, causing the microbial particles in the airflow to accelerate and impact on the surface of the petri dish containing nutrient agar medium, forming colonies visible to the naked eye after cultivation, counting the colonies and calculating the concentration of living individuals in the air according to the The colonies are counted and the concentration of viable individuals in the air is calculated from the volume sampled.
1 Technical specifications
Table 1 Technical specifications
| Main parameters | Parameter range | Resolution | Accuracy |
| Sampling flow rate | (5 to 30)L/min | 0.1L/min | Better than ±2.5% |
| Sampling time | 1min~99h59min | 1min | Not more than ±0.2% |
| Pre-calculation pressure | (-30 to 0)kPa | 0.01kPa | Better than ±2.5% |
| Flow stability | Better than ±2.5% | ||
| Flow repeatability | Better than ±2.0% | ||
| Load capacity | Greater than 20kPa (at 28.3L/min flow) | ||
| Battery operating time | Greater than 8 hours | ||
| Capture rate | ≥98% | ||
| Class VI Anderson sampling head (standard) | First stage: >7 .0μm Pore size 1.18mm | ||
| Second stage: 4.7μm - 7.0μm Pore size 0.91mm | |||
| Tertiary: 3.3μm - 4.7μm Pore size 0.71mm | |||
| Level 4: 2.1μm - 3.3μm Pore size 0.53mm | |||
| Level 5: 1.1μm - 2.1μm Pore size 0.34mm | |||
| Grade 6: 0.65μm - 1.1μm Pore size 0.25mm | |||
| Sampling period | Any setting within 99 hours 59 minutes | ||
| Noise | ≤60db | ||
| Working power | 220V/AC or DC12V | ||
| Power | ≤ 45W | ||
| Host size | 183×196×225(mm) Approx. 3.3Kg | ||
| Sampling head size | 194mmH x 106mmD approx. 1.8Kg | ||
| Different sampling heads are available as an option |
Class II Anderson sampling head Class VI Anderson sampling head (standard) Class VIII Anderson sampling head | ||
2 Working conditions青岛国瑞英文版空气微生物采样器
1) Working voltage: 220V AC, 50HZ or 24V DC
2) Ambient temperature: (-20 to 60)°C.
3) Ambient humidity: (0 to 85)%RH.
4) Atmospheric pressure: (50 to 130) kPa.
5) Suitable environment: Non-explosion-proof applications.
6) When working in the field, measures should be taken to protect against rain, snow, dust and exposure to sunlight.
原创作者:青岛国瑞力恒环保科技有限公司